Z-1164-2026: Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infini

Recall record Z-1164-2026 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1164-2026
Company in FDA record
Elekta, Inc.
FDA classification date
Jan 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue and electrical grounding of systems.

Source facts

Action
On 12/14/2025, the firm emailed an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" Letter to customer informing them that Elekta has identified a safety issue relating to the modulator discharge rod, where the supplier has been fitting the crimp in the crimp tool back to front. The issue was discovered during production. Note - The discharge rod is used by Field Service Engineers (FSE) when completing service activities on the machine around the HT and RF (High Tension (high voltage) and Radio Frequency) assembly. Instructions to customer: No actions are required to be taken by the Clinical user. Instructions for Field Service Engineers: Possible impact on the Field Service Engineer:- The FSE uses the HT discharge rod to fully discharge the voltage capacitor. Once the FSE has completed the procedure using the discharge rod, the FSE will start to work on the system around the HT and RF assembly. If the discharge rod is not fully connected to the ground or not connected at all, the system will still be fully charged and the FSE could receive an electric shock. Elekta will release an Important Field Safety Modification (IFSM) to identify potentially faulty discharge rod crimping and provide further instructions. The IFSM will include procedures for verifying the reliability of the discharge rod crimping and replacing affected components if necessary. local Elekta service representative will contact you to arrange for this activity to be completed.
Initiated
2025-12-14
Root cause
Process control
Distribution
Worldwide
Recall number
Z-1164-2026
Recall status
Open, Classified
CDRH record ID
217320
FDA classified
2026-01-23
Recall event number
97958

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗