Open original FDA record ↗Recall record
- FDA identifier
- Z-1165-06
- Company in FDA record
- St Jude Medical CRMD
- FDA classification date
- Jun 24, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sterility-Some of the sterile pouches in which these products are packaged may not have been properly sealed during their packaging process.
Source facts
- Action
- The attached 'Dear Hospital Administrator' letter was sent out to our US sales representatives via FedEx on June 1, 2006. The sales representatives have been instructed to send back all of the devices in their inventory and to retrieve and return any remaining devices that were placed into hospital inventory. They are also instructed to deliver the attached 'Dear Hospital Administrator' letter to the appropriate person at each hospital where product is to be removed from the hospital shelf.
- Initiated
- 2006-06-01
- Root cause
- Other
- Terminated
- 2012-03-26
- Distribution
- Hospitals/Medical Centers in AZ, CA, MS, NY, & PA
- Recall number
- Z-1165-06
- Recall status
- Terminated
- CDRH record ID
- 46475
- FDA classified
- 2006-06-24
- Recall event number
- 35616