Z-1168-2020: Toco MP Transducer (M2734B), that may have been serviced with the following: Pacific Medical Philips M273

Recall record Z-1168-2020 for Pacific Medical Group Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1168-2020
Company in FDA record
Pacific Medical Group Inc.
FDA record creation date
Feb 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.

Source facts

Action
On 09/11/19, gent Medical Device Recall notices were emailed to customers. Customers were advised of the following possible failures associated with the post-market events: 1) Fluid damage/intrusion, solder cracks, component failure, 2) Damaged membrane or cracked case, 3) Internal short from usage, broken connector from misuse, 4) Broken leads, loose coil, 5) Loose crystal from failure of adhesive, internal short or improper grounding, 6) Miscalibration of the device/waveform, 7) Overheat of device. Malfunction of the device is likely to be recognized easily because the device will not function properly without intact components. If wave forms are not capturing the clinical data including fetal heart rate, oxygen saturation, patient's temperature or uterine activity, the device should not be used. Immediate actions to be taken by the Customer/User: - Inspect the device before each use to observe for any signs of device damage including cracked case, broken leads, loose coils, or damaged components and immediately remove the device from clinical service. - If any fluid intrusion is noted, remove the device from service immediately. - Any signs that the device is overheating should be recognized as a potential serious issue and should not be used and/or removed from the patient immediately. - If waveforms are not capturing on the monitor or there is artifact, this may be due to miscalibration of the device or a component malfunction. Remove the device from service. If there are no signs of product damage or deterioration, the serviced devices may continue to be used with continued monitoring of device performance. Maintain this notification at your facility and review the content of this letter with your staff. Complete and return the customer acknowledgement form to the recalling firm. If you have subsequently resold or distributed the affected product, please ensure that this information is provided to your customer(s). Customers with additio
Initiated
2019-09-11
Root cause
Device Design
Distribution
Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.
Recall number
Z-1168-2020
Recall status
Open, Classified
CDRH record ID
176765
FDA record created
2020-02-12
Recall event number
83610

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗