Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.
Source facts
Action
On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.
Initiated
2021-01-19
Root cause
Process change control
Distribution
US nationwide distribution.
Recall number
Z-1168-2021
Recall status
Open, Classified
CDRH record ID
185563
FDA record created
2021-03-04
Recall event number
87221