Z-1170-2025: Brand Name: Sertera Biopsy Device Product Name: Sertera Biopsy Device Model/Catalog Number: Sertera-14 So

Recall record Z-1170-2025 for Hologic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1170-2025
Company in FDA record
Hologic, Inc
FDA classification date
Feb 13, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement.

Source facts

Action
On January 15, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Our goal is to create as little disruption to your practice as possible. Below outlines the next steps for customers with affected product(s): " Please quarantine any product(s) from the impacted lots you currently have in your inventory. Impacted lots can be returned through IQVIA as part of the voluntary recall. Please note that you will be issued a credit on your account for the affected product(s). " Refer to the Customer Response Form Instructions to identify your product inventory that is subject to this action. " Confirm receipt of this communication by completing the online Customer Response Form. Please complete and submit the online Customer Response Form even if you no longer have the affected product in inventory. Hologic has partnered with IQVIA to conduct follow up communications should no response be received to this letter. " Hologic will issue a Return Goods Authorization upon receipt of the Customer Response Form and the return of all affected products to Hologic. " If you wish to reorder product, please place a new order via your preferred procurement method. We value your business and thank you for your cooperation. If you have any questions about this communication, please contact BreastHealth.Support@hologic.com or 1-877-371-4372.
Initiated
2025-01-15
Root cause
Process control
Distribution
Domestic: Nationwide Distribution
Recall number
Z-1170-2025
Recall status
Open, Classified
CDRH record ID
211921
FDA classified
2025-02-13
Recall event number
96075

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗