Open original FDA record ↗Recall record
- FDA identifier
- Z-1171-2016
- Company in FDA record
- Sekisui Diagnostics P.E.I. Inc.
- FDA record creation date
- Mar 14, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Crystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover
Source facts
- Action
- Sekisui sent an Important Field Correction - Customer Notification letter dated January 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Discard all reagent in house with a current expiration date of 2016-09. 2. Update the expiration on all remaining kits on hand with the new expiration dates provided 3. When adding additional reagent to the analyzer, a new reagent wedge should be used. 4. If crystals are observed within reagent stored on-board, reagent should be discarded and replaced with fresh reagent. 5. Complete, sign and return the attached CONFIRMATION OF NOTIFICATION form by FAX to the number provided. This will indicate receipt of this Field Correction, act as confirm action of compliance with the actions,and allow for issuance of replacement material for expired product on hand (lots 47677, 47694, 47695). Questions contact Technical Services at 1-800-565-0265 or via email at PEIDiagnosticTechnical@sekisui-dx.com
- Initiated
- 2016-01-22
- Root cause
- Device Design
- Terminated
- 2016-08-09
- Distribution
- Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.
- Recall number
- Z-1171-2016
- Recall status
- Terminated
- CDRH record ID
- 143121
- FDA record created
- 2016-03-14
- Recall event number
- 73129