Open original FDA record ↗Recall record
- FDA identifier
- Z-1171-2023
- Company in FDA record
- Summit Medical, LLC
- FDA classification date
- Feb 24, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sterile product pouches were not sealed.
Source facts
- Action
- The firm issued "URGENT RECALL NOTICE" Letters dated 1/13/2023 via email. The letter explained the issue and requested the consignee remove the product from inventory and inspect the pouch to verify it is sealed. If the pouch is not sealed, the consignee is to cut the product with scissors and dispose of it. Results of their inspection are to be reported to the recalling firm by email using the enclosed form. A number is listed to arrange for replacement product. The consignee is to indicate on the form that they have read and understand the recall instructions, have checked their stock and disposed of inventory affected, and that they have none of the product left in inventory. If you have any questions, contact Director of Quality Assurance at 651-789-3939 or email: jesselman@innoviamedical.com.
- Initiated
- 2023-01-13
- Root cause
- Process control
- Terminated
- 2026-08-14
- Distribution
- US Distribution was made to CO, FL, GA, LA, MA, and ME. There was no government/military distribution. OUS Foreign distribution was made to Canada.
- Recall number
- Z-1171-2023
- Recall status
- Terminated
- CDRH record ID
- 198119
- FDA classified
- 2023-02-24
- Recall event number
- 91527