Reason stated in the source
Poor welding may cause the device(s) to fail, which may result in the user, falling in the bathroom and/or a lack of assistance with bathing or toileting.
Source facts
- Action
- On 6/8/2007, a letter was sent to all dealers and each direct account customer/user followed by a phone call to ensure the notification letter was received. Customers were instructed to discontinue use of the devices pending receipt of modification kits/ instructions to be shipped priority overnight. The firm plans to call customers/dealers to support them with the installations and ensure smooth implementation. Dealers were instructed to conduct recall to the user level.
- Initiated
- 2007-06-01
- Root cause
- Device Design
- Terminated
- 2008-08-19
- Distribution
- Nationwide Distribution including states of PA, TX, NY, MI, AL, FL, CA, OH, CT, WA, MD, OR, NC, NH, VT, IL, CO, HI, LA, GA, TN, OK, MA, KS, SC, ME, GA, MN, IN, and DE.
- Recall number
- Z-1172-2007
- Recall status
- Terminated
- CDRH record ID
- 52939
- FDA classified
- 2007-08-17
- Recall event number
- 38096