A number of MDNC-3421 displays were shipped while not being DICOM-compliant. These displays will not pass DICOM acceptance tests.
Source facts
Action
On January 22, 2021 the firm notified the importer in the USA, Barco Inc.
Initiated
2021-01-22
Root cause
Process change control
Terminated
2022-11-18
Distribution
WI.
Recall number
Z-1172-2021
Recall status
Terminated
CDRH record ID
185729
FDA record created
2021-03-04
Recall event number
87288