Skip to content

Public FDA record

Z-1172-2021: Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY

Recall record Z-1172-2021 for Barco N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1172-2021
Company in FDA record
Barco N.V.
FDA record creation date
Mar 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A number of MDNC-3421 displays were shipped while not being DICOM-compliant. These displays will not pass DICOM acceptance tests.

Source facts

Action
On January 22, 2021 the firm notified the importer in the USA, Barco Inc.
Initiated
2021-01-22
Root cause
Process change control
Terminated
2022-11-18
Distribution
WI.
Recall number
Z-1172-2021
Recall status
Terminated
CDRH record ID
185729
FDA record created
2021-03-04
Recall event number
87288

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗