The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
Source facts
Action
On January 3, 2024, the firm notified customers via URGENT MEDICAL DEVICE CORRECTION letters.
Fujifilm will offer to correct the formula that calculates the LV Mass (2D Bullet) in the Synapse database. Until the update is scheduled and completed, customers may continue to safely use the Pediatric Workflow modules listed in the notice as long as the LV Mass (2D Bullet) equation is not the main factor for a diagnosis.
Customers should coordinate with the Fujifilm service engineer to schedule the correction of their device when contacted.
Please contact Fujifilm's field action coordinator, Kotei Aoki, via phone at 765-246-2931 or email at kotei.aoki@fujifilm.com, if you have any further questions regarding this field action.
Initiated
2024-01-03
Root cause
Software design
Distribution
US domestic distribution nationwide. International distribution to Israel.
Recall number
Z-1172-2024
Recall status
Open, Classified
CDRH record ID
205710
FDA classified
2024-02-21
Recall event number
93855
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.