- FDA identifier
- Z-1173-2007
- Company in FDA record
- Sendx Medical Inc
- FDA classification date
- Aug 18, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Error in the Ca12 calibration value for pCO2 recorded in the cal pack barcode for lot 17323. This error will result in falsely elevated pCO2 results at the high end of the measuring range.
Source facts
- Action
- A total of four (4) distributors were sent the recall package (Field Action Notice) by email on 6-29-07. The distributors were provided background information and specific instruction that all Cal packs from lot 17323 should be removed from customer sites and destroyed. The Recall package included a Management Cover for Field Action Notes, a Customer Letter and the Field Action Notice #915-261. All communications regarding field actions associated with this recall were distributed using the Radiometer Field Action Note (FAN) system.
- Initiated
- 2007-06-19
- Root cause
- Other
- Terminated
- 2011-11-07
- Distribution
- Worldwide Distribution --- USA including state of OH and countries of China, Denmark Latvia, Mexico & Spain.
- Recall number
- Z-1173-2007
- Recall status
- Terminated
- CDRH record ID
- 53481
- FDA classified
- 2007-08-18
- Recall event number
- 38359