Z-1173-2014: Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versi

Recall record Z-1173-2014 for Horiba Instruments, Inc dba Horiba Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1173-2014
Company in FDA record
Horiba Instruments, Inc dba Horiba Medical
FDA classification date
Mar 11, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

HORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.

Source facts

Action
Horiba sent an "URGENT SAFETY NOTICE letter datedFebruary 13, 2014 via by mail to customers. The letter identified the affected product, issue, impact, actions to be taken. Customers were instructed to share this information with laboratory staff and retain this notification as part of their laboratory documentation. Customers were requested to complete and return the enclosed response form via one of the following methods: Fax: 949-453-0600 Email: jim.knowles@horiba.com Mail: HORIBA Medical Attention: Jim Knowles, Recall Coordinator 34 Bunsen Drive Irvine, CA 92618. For questions contact your local Horiba Medical representative.
Initiated
2014-02-14
Root cause
Software design
Terminated
2014-08-22
Distribution
USA Nationwide Distribution
Recall number
Z-1173-2014
Recall status
Terminated
CDRH record ID
125675
FDA classified
2014-03-11
Recall event number
67528

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗