Skip to content

Public FDA record

Z-1173-2018: Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Rec

Recall record Z-1173-2018 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1173-2018
Company in FDA record
Biomet, Inc.
FDA classification date
Mar 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL letter will be issued to customers titled "Affected Product: Comprehensive Segmental Revision System (SRS)" instructing customers to do the following: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete the attached Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. ***Updated 3/31/2020*** On March 31, 2020, Zimmer Biomet expanded the recall to include two additional lots: Lot 434480 (for Item No. 211253) and Lot 434470 (for Item No. 211251). Customer instructions remain the same.
Initiated
2018-02-28
Root cause
Under Investigation by firm
Terminated
2020-08-11
Distribution
Distribution US Nationwide
Recall number
Z-1173-2018
Recall status
Terminated
CDRH record ID
162212
FDA classified
2018-03-23
Recall event number
79399

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗