Z-1173-2021: Sensis / Sensis Vibe systems with software version VD12 and equipped with HP Flex Pro-C PC. Diagnostic a

Recall record Z-1173-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1173-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Mar 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

System may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system

Source facts

Action
Siemens issued Urgent Medical Device Correction via Customer Safety Advisory Notice AX010/21/S dated 1/27/21distributed to all affected customers. Additionally, a solution to eliminate the root cause of this problem is being distributed via Update Instruction AX009/21/S to all affected customers. Letter states reason for recall, health risk and action to take: The freeze is a total lock-up of the system and requires a power cycle of the PC to recover. Please make sure that an alternative system can be used to continue treatment in such cases.Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Initiated
2021-01-27
Root cause
Software design
Terminated
2021-10-25
Distribution
US Nationwide distribution in the states of AK, AR, CA, GU, IA, IL, MI, MO, NC, NE, OR, PA, PA, PR, SD, TN, TX, WA.
Recall number
Z-1173-2021
Recall status
Terminated
CDRH record ID
185684
FDA record created
2021-03-04
Recall event number
87273

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗