Open original FDA record ↗Recall record
- FDA identifier
- Z-1173-2024
- Company in FDA record
- IMPULSE DYNAMICS (USA) INC
- FDA classification date
- Feb 22, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.
Source facts
- Action
- On January 2, 2024, the firm sent letters to field representatives to be given to affected physicians. Summary of the safety communication can be found online: https://impulse-dynamics.com/safety-information/ The firm recommends all patients with implant card showing model CCM X11, who have experienced the "A9" charge error, charge their implanted device to no more than 75% of full capacity to circumvent the issue. See link above for instructions. If a patient encounters an A9 error code, CCM therapy has been suspended and must be manually reset. Patients should schedule time with an Impulse Dynamics representative or their physician to manually reset their device. The firm is working on a software update to address this issue. Patients and clinicians may determine if a specific device is affected by providing the serial number of any specific device to the Technical Support Hotline at 866-312-5370.
- Initiated
- 2024-01-05
- Root cause
- Software design
- Distribution
- Domestic US distribution nationwide. International distribution to Austria Croatia Czech Republic Germany Italy Macau Poland Romania Russia Spain Sweden.
- Recall number
- Z-1173-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205602
- FDA classified
- 2024-02-22
- Recall event number
- 93791