Skip to content

Public FDA record

Z-1173-2024: OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite

Recall record Z-1173-2024 for IMPULSE DYNAMICS (USA) INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1173-2024
Company in FDA record
IMPULSE DYNAMICS (USA) INC
FDA classification date
Feb 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.

Source facts

Action
On January 2, 2024, the firm sent letters to field representatives to be given to affected physicians. Summary of the safety communication can be found online: https://impulse-dynamics.com/safety-information/ The firm recommends all patients with implant card showing model CCM X11, who have experienced the "A9" charge error, charge their implanted device to no more than 75% of full capacity to circumvent the issue. See link above for instructions. If a patient encounters an A9 error code, CCM therapy has been suspended and must be manually reset. Patients should schedule time with an Impulse Dynamics representative or their physician to manually reset their device. The firm is working on a software update to address this issue. Patients and clinicians may determine if a specific device is affected by providing the serial number of any specific device to the Technical Support Hotline at 866-312-5370.
Initiated
2024-01-05
Root cause
Software design
Distribution
Domestic US distribution nationwide. International distribution to Austria Croatia Czech Republic Germany Italy Macau Poland Romania Russia Spain Sweden.
Recall number
Z-1173-2024
Recall status
Open, Classified
CDRH record ID
205602
FDA classified
2024-02-22
Recall event number
93791

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗