Open original FDA record ↗Recall record
- FDA identifier
- Z-1176-2009
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Apr 9, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.
Source facts
- Action
- Consignees were notified by an Urgent Medical Device Recall letter, dated 03/03/2009, via UPS Next Day. Hospitals were instructed to immediately discontinue use of the products, inspect the scissors for the date codes and to return the product to Teleflex Medical for replacement. Distributors were instructed to forward the letter to their consignees to retrieve relevant product from those locations. A Recall Acknowledgement Form was attached to be completed and faxed back to Teleflex Medical. A second mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or they may be contacted by phone, e-mail or fax. A record of these contacts will be maintained. Affected product retrieved by Teleflex Medical will be destroyed and replacement product provided to the consignee. Records of destruction will be maintained.
- Initiated
- 2009-03-03
- Root cause
- Process control
- Terminated
- 2011-03-31
- Distribution
- Worldwide Distribution -- USA including states of AL, AZ, CA, CO, DC, DE, FL, IA, IN, KS, MA, MD, MI, MO, MS, NH, NJ, NY, NV, OH, OK, OR, PA, RI, TN, TX, UT, and WA, and countries of Australia, Canada, France, Germany, India, Japan, and Singapore.
- Recall number
- Z-1176-2009
- Recall status
- Terminated
- CDRH record ID
- 79683
- FDA classified
- 2009-04-09
- Recall event number
- 51300