Open original FDA record ↗Recall record
- FDA identifier
- Z-1177-2012
- Company in FDA record
- Pelton & Crane Company
- FDA classification date
- Mar 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Modification of the Pelton & Crane Spirit 3000 unit disabled the TLC SAFEdrive feature rendering it ineffective in detecting whether the handpiece is potentially damaged or heating up.
Source facts
- Action
- Pelton & Crane sent an Urgent Medical Device Recall letter dated February 20, 2012 to all affected customers. The letter identifies the affected product, problem and actions to be taken. The letter states that a Pelton & Crane representative will be contacting the customers to schedule a service appointment with the end user of the device so that the affected TLC SAFEdrive circuit board can be replaced and returned to Pelton & Crane. Customers were instructed to complete the attached Recall Acknowledgement/Return Form and fax it to Pelton & Crane, attention Frank Ray, at (704) 587-7250. The letter also requested customers to mail the original Recall Acknowledgement/Return Form in the enclosed self addressed stamped envelope.
- Initiated
- 2012-02-20
- Root cause
- Employee error
- Terminated
- 2012-03-28
- Distribution
- (USA) Nationwide Distribution
- Recall number
- Z-1177-2012
- Recall status
- Terminated
- CDRH record ID
- 107682
- FDA classified
- 2012-03-06
- Recall event number
- 61262