Z-1177-2021: VentStar Anesthesia (N) 180 - Drager Breathing Circuits/Anesthesia Sets Order Number: MP00333 - Product U

Recall record Z-1177-2021 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1177-2021
Company in FDA record
Draeger Medical, Inc.
FDA record creation date
Mar 5, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Y-piece may become detached from the ventilation hose either before or during the ventilation process, a leak will be caused that can hinder ventilation of the patient.

Source facts

Action
Draeger Inc. (US) issued Urgent Medical Device Recall Letter via Certified Mail on January 29, 2021. The letter states reason for recall, health risk and action to take: Please check the quality of the adhesive joint before use by holding the Y-piece with one hand and then pulling (not twisting) each of the two hoses one after another so that the hose just starts to stretch visibly. Hoses that you have tested in this manner may be used if they pass the leakage test before use. Only use the hoses with devices that we have declared compatible in the Instructions for Use or accessories list of the basic device. Ensure that the patient is adequately monitored and select suitable alarm limits. During installation and operation, make sure to avoid any unnecessary tensile forces, and do not use the hoses for longer than is specified in the Instructions for Use. Please inspect your existing stock before use by following the above inspection instructions. Please also complete and return the attached Customer Reply and Order form to confirm to us that you have received this information. If you wish, we will be happy to replace any of your unused stocks that were manufactured between 01.01.2018 and 13.11.2020. The form may also be used to order Free of Charge replacements. Questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Initiated
2021-01-29
Root cause
Process control
Terminated
2021-11-01
Distribution
US Nationwide distribution.
Recall number
Z-1177-2021
Recall status
Terminated
CDRH record ID
185673
FDA record created
2021-03-05
Recall event number
87267

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗