Reason stated in the source
It was discovered in some instances the detector did not reset as quickly as the user prompted a change form one image receptor (auxiliary) to the next, and then back again.
Source facts
- Action
- An Important Electronic Product Radiation Warning letter, dated February 28, 2012, was sent to customers. The letter identified the affected product, a description of the issue, the hazard involved, along with the actions to be taken and planned. Customers were to use one of the three work-a-rounds provided. The issue will be corrected with Field Change Order 70600064, which consists of a Software Update 4.0.2SP Service Pack installed on the system. If further information or support is needed, customers can contact Dominic Siewko at 978-659-7936.
- Initiated
- 2011-04-01
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2016-08-23
- Distribution
- Worldwide Distribution - Including USA.