Open original FDA record ↗Recall record
- FDA identifier
- Z-1178-2013
- Company in FDA record
- Alere San Diego, Inc.
- FDA classification date
- Apr 25, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Alere Cholestech LDX hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the same patient sample.
Source facts
- Action
- Alere sent an Urgent Medical Device Recall letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to block all stock of the affected lot in their warehouses; discontinue use and discard all stock on hand, complete the provided Customer Verification form to document disposal, contact their primary consignees and advise them of the required discontinuation and disposal of the affected lot. For question call 877-308-8289.
- Initiated
- 2013-03-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-03-12
- Distribution
- USA Nationwide Distribution
- Recall number
- Z-1178-2013
- Recall status
- Terminated
- CDRH record ID
- 116831
- FDA classified
- 2013-04-25
- Recall event number
- 64701