Open original FDA record ↗Recall record
- FDA identifier
- Z-1179-2013
- Company in FDA record
- Chicago X-Ray Systems, Inc.
- FDA classification date
- May 21, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The IntraOs 70 system was found to have missing or incomplete system labels as required by 21 CFR 1010.2 & 1010.3. Also, the user's manual was found to have missing or improperly labeled information as required by 21 CFR 1020.30(h).
Source facts
- Action
- FONA will have upgraded copies of the manuals that will be made available on the firms website for free download and may also be mailed to each customer or end user should they request it. They will also provide users and distributers with corrected system labels to be placed on the x-ray control, beam limiting device, and tube housing assembly along with instructions for the proper application of these labels. They will provide users and distributers with an amended "Technical Data" page for insertion int the user's manual. Further questions please call (847) 459-3889.
- Initiated
- 2012-07-03
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-09-05
- Distribution
- Worldwide Distribution- USA (nationwide) and Latin America.
- Recall number
- Z-1179-2013
- Recall status
- Terminated
- CDRH record ID
- 117675
- FDA classified
- 2013-05-21
- Recall event number
- 65018