Open original FDA record ↗Recall record
- FDA identifier
- Z-1179-2018
- Company in FDA record
- Keystone Dental Inc
- FDA record creation date
- Mar 24, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
A locking plastic abutment sleeve, used by dental technicians in dental laboratories to aid in the creation of the final restorative crown, was packed as a non-locking plastic abutment sleeve.
Source facts
- Action
- The firm, Keystone Dental, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 10/24/217 to its consignee. The letter described the product, problem and actions to be taken. The consignee was instructed to do the following: Return any product remaining in inventory to Keystone Dental and replacement product will be sent to you. Note: These lots were the only affected product and all other inventory you may have can be used. Complete and return the Customer Relations Fax Back Form to Customer Relations (1-866-903-9272-US only/ 1-781-328-3400). If you have any questions, please contact Customer Relations Manager at: Toll Free 866-902-9272; Fax 866-903-9272 and CustomerSupport@KeystoneDental.com.
- Initiated
- 2017-10-16
- Root cause
- Labeling mix-ups
- Terminated
- 2019-04-17
- Distribution
- International Distribution to: Germany only.
- Recall number
- Z-1179-2018
- Recall status
- Terminated
- CDRH record ID
- 161839
- FDA record created
- 2018-03-24
- Recall event number
- 79297