Skip to content

Public FDA record

Z-1179-2018: STAGE-1 RDS COC Abutment Plastic Sleeve, Non-Lock, Multi-Unit; Catalog Number: S2437-01-1K

Recall record Z-1179-2018 for Keystone Dental Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1179-2018
Company in FDA record
Keystone Dental Inc
FDA record creation date
Mar 24, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

A locking plastic abutment sleeve, used by dental technicians in dental laboratories to aid in the creation of the final restorative crown, was packed as a non-locking plastic abutment sleeve.

Source facts

Action
The firm, Keystone Dental, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 10/24/217 to its consignee. The letter described the product, problem and actions to be taken. The consignee was instructed to do the following: Return any product remaining in inventory to Keystone Dental and replacement product will be sent to you. Note: These lots were the only affected product and all other inventory you may have can be used. Complete and return the Customer Relations Fax Back Form to Customer Relations (1-866-903-9272-US only/ 1-781-328-3400). If you have any questions, please contact Customer Relations Manager at: Toll Free 866-902-9272; Fax 866-903-9272 and CustomerSupport@KeystoneDental.com.
Initiated
2017-10-16
Root cause
Labeling mix-ups
Terminated
2019-04-17
Distribution
International Distribution to: Germany only.
Recall number
Z-1179-2018
Recall status
Terminated
CDRH record ID
161839
FDA record created
2018-03-24
Recall event number
79297

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗