Risk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of a harder rubber formulation and may be more difficult to penetrate.
Source facts
Action
On or about 6/15/2007, the firm sent advisory notice letters to the end users and the instructions for use have been updated by adding a precaution statement regarding the use of thick rubber stoppers. The firm is also updating the customer training program to include how to handle drug vials with thick concave rubber stoppers.
Initiated
2007-06-26
Root cause
Other
Terminated
2013-03-05
Distribution
Product was distributed throughout the US and Canada
Recall number
Z-1180-2007
Recall status
Terminated
CDRH record ID
53277
FDA classified
2007-08-22
Recall event number
38251
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.