Open original FDA record ↗Recall record
- FDA identifier
- Z-1181-2008
- Company in FDA record
- Hospira Inc.
- FDA classification date
- Aug 6, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect dosage labels created: Customers have experienced independent changes in drug concentration (mg/mL to mcg/mL and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.
Source facts
- Action
- Hospira sent letters dated 1/3/08 to the immediate recipients of the product via FedEx overnight letter on the same date. The letters informed the accounts of the reports of independent changes in drug concentration while editing a saved formulation using the Hospira LifeCare PCA Label Utility Software CD that was provided to some customers in a Hospira MedNet Drug Library Creation Team Introductory Kit-LifeCare PCA kit. The accounts were requested to remove the utility from any PC in their facility on which it is currently installed, destroy the CD and sign and return the enclosed removal verification form verifying that they have removed the software from their system. The letter listed commercially available software packages that the account could validate and implement in their facility. Any questions concerning the utility software withdrawal were directed to Hospira at 1-800-241-4002, option 4, then option 1.
- Initiated
- 2008-01-03
- Root cause
- Software design
- Terminated
- 2010-03-09
- Distribution
- Nationwide Distribution and Canada
- Recall number
- Z-1181-2008
- Recall status
- Terminated
- CDRH record ID
- 68332
- FDA classified
- 2008-08-06
- Recall event number
- 46846