Skip to content

Public FDA record

Z-1181-2021: Anesthesia gas-machine - Product Usage: intended for use in administering anesthesia while controlling th

Recall record Z-1181-2021 for Getinge Group Logistics America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1181-2021
Company in FDA record
Getinge Group Logistics America, LLC
FDA record creation date
Mar 5, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.

Source facts

Action
On 2/9/2021 Maquet Critical Care AB / Getinge issued an Urgent Medical Device Correction letter via FedEx for the Flow i anesthesia machine due to the potential for high FiCO2 values and/or a System Check Out failure due to an incomplete docking of the CO2 absorber to the patient cassette. The patients re-breathing without CO2 reduction could occur if the absorber is not sufficiently docked.
Initiated
2021-01-29
Root cause
Device Design
Terminated
2023-09-28
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, MA, MI, MO, NJ, NY, OH, OR, PA, TX, VA, WA, WV.
Recall number
Z-1181-2021
Recall status
Terminated
CDRH record ID
185764
FDA record created
2021-03-05
Recall event number
87301

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗