Open original FDA record ↗Recall record
- FDA identifier
- Z-1182-06
- Company in FDA record
- MDS Canada Inc. DBA MDS Nordion
- FDA classification date
- Jun 27, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions.
Source facts
- Action
- A User Bulletin (CUB-2004-01dated 7/5/04) was issued to affected users on 8/24/04 and 8/31/04. The bulletin provided 2 options for corrective action.
- Initiated
- 2004-07-08
- Root cause
- Other
- Terminated
- 2006-06-27
- Distribution
- The products were shipped to medical facilities in FL and KS. The product was also shipped to Kuwait, Spain, Honduras, Romania, France, Grrece, Bangladesh, Germany, Nepal, Canada, England, Thailand, Phillippines, Mexico, Columbia, India, Switzerland, Syria, Trinidad & Tobago, Ireland, Turkey, Algeria, Portugal, Poland, Iran, China, Cuba, Belgium, Hungary, Egypt, Italy, Vietnam, Sri Lanka, and Costa Rica.
- Recall number
- Z-1182-06
- Recall status
- Terminated
- CDRH record ID
- 46577
- FDA classified
- 2006-06-27
- Recall event number
- 35617