- Company in FDA record
- Philips Medical Systems
- FDA classification date
- Sep 5, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Patient Assignment: the information center will in most cases be associated with the "incorrect" patient. This occurs on both standalone and networked systems running fixed mode monitoring. (This does not occur on systems that run flex mode monitoring)
Source facts
- Action
- Philips Medical Systems issued a letter notification dated July 2007 " Urgent Device Correction". Customers with networked information centers will be corrected through a database server. Stand-Alone units will have a software correction. Software version J.00.25 will correct the problem. Users are advised not to use th 12 Lead Analysis and Export Option in fixed or flex mode monitoring unitl the software correction is made.
- Initiated
- 2007-07-16
- Root cause
- Other
- Terminated
- 2016-08-01
- Distribution
- Worldwide Distribution: USA including states of NJ, NY, OK, MI, and UT and Foreign: Australia, Finland, France, Netherlands, and the UK
- Recall number
- Z-1182-2007
- Recall status
- Terminated
- CDRH record ID
- 53538
- FDA classified
- 2007-09-05
- Recall event number
- 38400