Z-1182-2024: IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft

Recall record Z-1182-2024 for Alphatec Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1182-2024
Company in FDA record
Alphatec Spine, Inc.
FDA classification date
Feb 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to reports of intraoperative graft bolt implantation breakages.

Source facts

Action
On 01/26/2024, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customer informing them that, The Surgical Technique Guide, LIT-85141, and Instructions for Use, INS-130, were revised to require hole preparation prior to insertion of an IdentiTi SA Graft Bolt, as well as to add a new caution statement, to confirm with fluoroscopy that the graft bolt insertion angle is as close as possible to a 40-degree trajectory. The revised labeling mitigates potential fracturing of the Graft Bolt intra-operatively or post-operatively. Customer are instructed to: Please share this notice with all of those who need to be aware within your organization or any organization where the affected product has been transferred. - Inform doctors and surgical staff of the labeling (surgical technique and instructions for use) - Please fill out the last page of this letter to confirm that you have read this notification and will complete the necessary actions as described in this notification. - If you have any questions/concerns or require assistance with the correction, please contact by e-mailing QMSCompliance@atecspine.com or calling 1-800-922-1356.
Initiated
2024-01-26
Root cause
Device Design
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KS, KY, MD, MI, MN, MS, NC, NH, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, WA and WI. The country of New Zealand.
Recall number
Z-1182-2024
Recall status
Open, Classified
CDRH record ID
205817
FDA classified
2024-02-23
Recall event number
93898

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗