Open original FDA record ↗Recall record
- FDA identifier
- Z-1182-2024
- Company in FDA record
- Alphatec Spine, Inc.
- FDA classification date
- Feb 23, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to reports of intraoperative graft bolt implantation breakages.
Source facts
- Action
- On 01/26/2024, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customer informing them that, The Surgical Technique Guide, LIT-85141, and Instructions for Use, INS-130, were revised to require hole preparation prior to insertion of an IdentiTi SA Graft Bolt, as well as to add a new caution statement, to confirm with fluoroscopy that the graft bolt insertion angle is as close as possible to a 40-degree trajectory. The revised labeling mitigates potential fracturing of the Graft Bolt intra-operatively or post-operatively. Customer are instructed to: Please share this notice with all of those who need to be aware within your organization or any organization where the affected product has been transferred. - Inform doctors and surgical staff of the labeling (surgical technique and instructions for use) - Please fill out the last page of this letter to confirm that you have read this notification and will complete the necessary actions as described in this notification. - If you have any questions/concerns or require assistance with the correction, please contact by e-mailing QMSCompliance@atecspine.com or calling 1-800-922-1356.
- Initiated
- 2024-01-26
- Root cause
- Device Design
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KS, KY, MD, MI, MN, MS, NC, NH, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, WA and WI. The country of New Zealand.
- Recall number
- Z-1182-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205817
- FDA classified
- 2024-02-23
- Recall event number
- 93898