Open original FDA record ↗Recall record
- FDA identifier
- Z-1183-2008
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Aug 20, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Source facts
- Action
- Beckman Coulter mailed a Product Corrective Action (PCA) letter on October 17, 2007 to all Access and Access 2 lmmunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System customers. The customers were informed that during manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. The letter referenced a proactive initiative to inspect the power supply sled assemblies on all instruments referenced. A response form was included.
- Initiated
- 2007-10-17
- Root cause
- Process control
- Terminated
- 2012-04-27
- Distribution
- Nationwide Distribution and Canada
- Recall number
- Z-1183-2008
- Recall status
- Terminated
- CDRH record ID
- 67172
- FDA classified
- 2008-08-20
- Recall event number
- 46316