The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.
Source facts
Action
Quidel sent and Urgent Product Recall letter dated December 15, 2017.
The letter instructs the user to discontinue use of the device until new firmware can be installed. For further questions, please call (800) 874-1517.
Initiated
2017-12-15
Root cause
Software Manufacturing/Software Deployment
Terminated
2019-10-21
Distribution
US Distribution
Recall number
Z-1183-2018
Recall status
Terminated
CDRH record ID
161800
FDA record created
2018-03-26
Recall event number
79271