Z-1184-2008: Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modif

Recall record Z-1184-2008 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1184-2008
Company in FDA record
Beckman Coulter Inc
FDA classification date
Aug 20, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range.

Source facts

Action
Beckman Coulter mailed a Product Corrective Action (PCA) letter the week of October 8, 2007 to the customers who have the affected UniCel Dxl 800 Access lmmunoassay systems advising that there is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range. The letter further advises that Beckman Coulter will contact the customer to arrange a date to remove the modificatlons but until the modification is completed, the system should be operated in a room temperature of 20C to 30C. This can be acomplished by monitoring and recording room tamperature in the area of the instrument. Customers are also asked to review historical quality control records to ensure assay performance is within their laboratory specification and that patient results should be reviewed in context of other diagnostics tests and the clinical presentations of the patient. A response form is included.
Initiated
2007-10-08
Root cause
Device Design
Terminated
2012-01-11
Distribution
Nationwide and Canada
Recall number
Z-1184-2008
Recall status
Terminated
CDRH record ID
67204
FDA classified
2008-08-20
Recall event number
46336

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗