Reason stated in the source
This correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard.
Source facts
- Action
- The firm, SIEMENS, issued a "Customer Information" letter to its customers. The letter described the product, problem and actions to be taken. Siemens developed a new steel knob that complies with the latest standards; Siemens checked all ModuLeaf mMLC trolleys and if necessary refitted with the safety label and also arranged for an inspection and modification of the customers system- a Release of the safety update UI TH 026/11/S to distribute and hardware update commenced on February 7, 2007 by certified mail. If you have any questions, contact the Regulatory Specialist II at 925-602-8175.
- Initiated
- 2012-02-07
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-08-14
- Distribution
- Worldwide distribution: USA (nationwide) including states of: CO, IA, KY, NJ, OH and TX; and countries of: Algeria, Austria, Belgium, Brazil, China, Egypt, France, Germany, Hungary, India, Italy, Japan, Kuwait, Malaysia, Mexico, P.R., Russian Fed., Saudi Arabia, South Africa, Spain, Turkey, and United Kingdom.
- Recall number
- Z-1184-2012
- Recall status
- Terminated
- CDRH record ID
- 107631
- FDA classified
- 2012-03-07
- Recall event number
- 61237