Skip to content

Public FDA record

Z-1184-2024: Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 0

Recall record Z-1184-2024 for Medagent Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1184-2024
Company in FDA record
Medagent Inc
FDA classification date
Feb 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.

Source facts

Action
Rebstock issued letter to Distributors via email on 12-21-23. Letter states reason for recall, health risk and action to take: Please check your stock immediately. If you have holding arms affected by the recall in stock, they must be blocked immediately. 2. If you have resold the parts, instruct your customers not to reuse them under any circumstances. Ask your customers to return these parts immediately. 3. You will receive a confirmation form with this letter, please fill it out completely, sign it and send it back to us after receiving this information. If you do not have any affected products, please complete the confirmation form anyway and fax it to +49 (0)7424-98230 50 or email it to Eva.Stumpp@rebstock.de. 4. Return all affected holding arms to us. After receiving the goods, we will clarify the further procedure with you. 5. Forward this information to all employees in your facility who need to be informed. 6. Please also check whether you need to make a report in your country and let us know what further information you need from us: If you have any questions, please contact our PRRCs: Mr. Matthias Tribl Mrs. Eva Stumpp E-Mail: Matthias.Tribl@rebstock.de E-Mail: Eva.Stumpp@rebstock.de Phone.: +49 7424 98230-38 Phone.: +49 7424 98230-32
Initiated
2023-12-21
Root cause
Device Design
Distribution
TX Foreign: Algeria China Dominican Republic Germany India Japan Philippines UK
Recall number
Z-1184-2024
Recall status
Open, Classified
CDRH record ID
205620
FDA classified
2024-02-23
Recall event number
93840

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗