Open original FDA record ↗Recall record
- FDA identifier
- Z-1185-2021
- Company in FDA record
- Brainlab AG
- FDA record creation date
- Mar 8, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.
Source facts
- Action
- The Field Safety Notice / Product Notification Letter CAPA-20210210-002387 was distributed via email on 03/01/2021. An active reply by the consignees will be followed, if necessary via further different communication media. For future customers/facilities/systems (new installations or updates/upgrades): As of February 23, 2021: - the Field Safety Notice / product notification information is handed over to the customer during installation (before clinical use) by the according Brainlab Service representative, - a confirmation is required that the contained User Corrective Actions will be implemented before clinical use by the facility (to be adhered to by the users).
- Initiated
- 2021-03-01
- Root cause
- Software design
- Terminated
- 2023-03-20
- Distribution
- USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.
- Recall number
- Z-1185-2021
- Recall status
- Terminated
- CDRH record ID
- 185958
- FDA record created
- 2021-03-08
- Recall event number
- 87395