Open original FDA record ↗Recall record
- FDA identifier
- Z-1185-2026
- Company in FDA record
- NIPRO Technical Services, Inc.
- FDA classification date
- Jan 27, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
Source facts
- Action
- On 12/12/2025, the firm mailed/emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that specific lots of Conductivity Standard Solutions have been identified as not meeting specification requirements which potentially could result with inaccurate calibration leading to electrolyte imbalances consistent with hypo- or hypernatremia. Customer are instructed to" 1. Immediately stop using the affected products 2.Segregate and quarantine all affected units 3. Return the affected product for replacement or dispose of the product For questions associated with this recall or if further assistance is needed, contact Customer Service at 800-923-2486 or email sales@niprotechnical.com.
- Initiated
- 2025-12-12
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
- Recall number
- Z-1185-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217459
- FDA classified
- 2026-01-27
- Recall event number
- 98030