Skip to content

Public FDA record

Z-1185-2026: Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 1004

Recall record Z-1185-2026 for NIPRO Technical Services, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1185-2026
Company in FDA record
NIPRO Technical Services, Inc.
FDA classification date
Jan 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.

Source facts

Action
On 12/12/2025, the firm mailed/emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that specific lots of Conductivity Standard Solutions have been identified as not meeting specification requirements which potentially could result with inaccurate calibration leading to electrolyte imbalances consistent with hypo- or hypernatremia. Customer are instructed to" 1. Immediately stop using the affected products 2.Segregate and quarantine all affected units 3. Return the affected product for replacement or dispose of the product For questions associated with this recall or if further assistance is needed, contact Customer Service at 800-923-2486 or email sales@niprotechnical.com.
Initiated
2025-12-12
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Recall number
Z-1185-2026
Recall status
Open, Classified
CDRH record ID
217459
FDA classified
2026-01-27
Recall event number
98030

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗