Open original FDA record ↗Recall record
- FDA identifier
- Z-1187-2012
- Company in FDA record
- Integra LifeSciences Corp
- FDA classification date
- Mar 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Material used in the superior region of the InterFix CT Adapter (CTA) has a different density than that of the Treatment Table Couch Adapter (TTA). Resulting dosimetry can be affected if posterior fields are used. Integra estimates that a target dose decrease of 1.2% would result from a typical 360 degree TomoTherapy arc.
Source facts
- Action
- Integra sent a Product Technical Bulletin dated January 23, 2012, to all affected consignees via FedEx, email or telephone conference calls. The revised Operator Manual was provided with a Technical Bulletin to all consignees that have been shipped this system. Consignees were asked to distribute the information in the bulletin to members of the Radiotherapy Planning Team within their institution. Customers were also asked to sign and return the enclosed Product Technical Bulletin Acknowledgement Form. For questions regarding this recall call 888-772-7378.
- Initiated
- 2012-01-23
- Root cause
- Device Design
- Terminated
- 2012-10-01
- Distribution
- Worldwide Distribution - USA including DC, IN, FL, HI, LA, NY, MD, MI, OH,OR, WI, WY, Foreign: Great Britain, Italy, and France
- Recall number
- Z-1187-2012
- Recall status
- Terminated
- CDRH record ID
- 107363
- FDA classified
- 2012-03-08
- Recall event number
- 61131