FDA identifier
Z-1187-2021
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Mar 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.

Source facts

Action
On 7/24/2012 the firm sent recall notificaiton letters to their customers.
Initiated
2012-02-23
Root cause
Nonconforming Material/Component
Terminated
2021-07-08
Distribution
US Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.
Recall number
Z-1187-2021
Recall status
Terminated
CDRH record ID
152991
FDA record created
2021-03-08
Recall event number
76372

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗