Open original FDA record ↗Recall record
- FDA identifier
- Z-1188-2008
- Company in FDA record
- Kentec Medical Inc
- FDA classification date
- Jun 26, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incompatible components: The Finger Clip Sensors - with a Serial Number prefix of PA514, PA515, YE508, ZC502 and ZG512 - are missing an additional diode and higher resistor in its existing configuration making them inoperable when used to interface with those Ohmeda Tuffsat monitors previously re-configured by Ohmeda to include a detection circuit for diodes.
Source facts
- Action
- Consignees were notified via telephone on 5/18/07 and an "Important Voluntary Market Withdrawal Notice" letter on 5/23/07. The letter instructed consignees to segregate and return all affected devices. The letter included an "Acknowledgement of Market Withdrawal Notification" form for customer to complete and return to the recalling firm. For additional information, contact 1-800-825-5996.
- Initiated
- 2007-05-18
- Root cause
- Process change control
- Terminated
- 2012-02-29
- Distribution
- Nationwide Distribution including states of AZ, CA, CO, FL, KY, LA, MA, NC, NJ, TX and UT.
- Recall number
- Z-1188-2008
- Recall status
- Terminated
- CDRH record ID
- 67366
- FDA classified
- 2008-06-26
- Recall event number
- 46397