Z-1188-2016: The Vesa 75/100 adaptation with rotation is an accessory to a surgical camera intended to be used to reco

Recall record Z-1188-2016 for Richard Wolf Medical Instruments Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1188-2016
Company in FDA record
Richard Wolf Medical Instruments Corp.
FDA record creation date
Jul 9, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Supplier of the "VESA 75 / 100 adaptation with rotation", which is part of their monitor mounts and support arms may be missing riveting.

Source facts

Action
RICHARD WOLF sent an Urgent Safety Information letter dated March 2, 2016, to all affected consignees. The letter described the nature of the recall regarding the monitor holding plate (VESA 75/100 mount) in our Flexion-Port 300, 350 and 400 swivel arm product series. According to the firm insufficient riveting of the rotation elements the rivets may loosen during continuous use (e.g. when adjusting the tilt angle) and in an extremely unlikely event the monitor holding plate (VESA 75/100 mount) may loosen. According to Richard Wolf this can result in a potential danger for the patient, user and others which we want to prevent with this measure. The firm is asking all end users to take the following step: Discontinue use of the affected devices until the conversion kits have been installed by a qualified service technician. Make sure that all users of affected medical devices are informed in their organization of the urgent safety information. If the end user has passed on products to third parties they should pass on a copy of this information and inform the recall contact person at Richard Wolf. Please make sure that the information is observed at their organization until all necessary internal measures have been completed. The end user should provide confirmation of receipt of this urgent safety information and answer the enclosed questionaire form which the firm asks to send to Richard Wolf at the following e-mail, marek.rast@richard-wolf.com by February 22, 2016. The firm asking that recipients to acknowledge receipt of the safety information in order to avoid receiving more reminders from Richard Wolf. The firm will send new monitor holding plate(s}, (643301264 conversion kit for monitor holding plate) once the form is received. The firm asks that qualified service technician install the new monitor holding plate (643301264 Conversion kit for monitor holding plate) in accordance with the enclosed assembly instructions and discard
Initiated
2016-02-29
Root cause
Process control
Terminated
2019-07-03
Distribution
Nationwide Distribution including Virginia, Pennsylvania, Florida, California, Texas, North Carolina and Illinois.
Recall number
Z-1188-2016
Recall status
Terminated
CDRH record ID
144067
FDA record created
2019-07-09
Recall event number
73348

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗