Skip to content

Public FDA record

Z-1188-2023: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0

Recall record Z-1188-2023 for Maquet Medical Systems USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1188-2023
Company in FDA record
Maquet Medical Systems USA
FDA classification date
Feb 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.

Source facts

Action
On February 13, 2023, the firm notified customers via Urgent Medical Device Correction letters. Customers were informed of the potential packaging issues and the associated risks to health. Customers were instructed to stop use and return all unused and unexpired affected product for credit. If customers have no alternative product, they may choose to continue use of the product, but they must perform a visual inspection of the primary packaging, checking for visible stress marks or damage to the packaging. If there are visible stress marks in the packaging, the customer should not use the product. Please note that there is a possibility that packaging defects may not be detectable by the naked eye. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 9GETUSA / (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2023-02-13
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czechia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritius, Myanmar, Mexico, Moldova, Morocco, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of North Macedonia, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, The Republic of Korea, The Russian Federation, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Vietnam.
Recall number
Z-1188-2023
Recall status
Open, Classified
CDRH record ID
198190
FDA classified
2023-02-28
Recall event number
91569

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.