Skip to content

Public FDA record

Z-1189-2010: AOS Trochanteric Nail, 12mm x 17cm x 130, Product Part Number: 1041-170.

Recall record Z-1189-2010 for Advanced Orthopaedic Solutions Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1189-2010
Company in FDA record
Advanced Orthopaedic Solutions Inc
FDA classification date
Mar 22, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated after Advanced Orthopaedic Solutions (AOS) became aware of a manufacturing defect involving the short Trochanteric Nail, where one piece was found to be incorrectly bent. If this issue were to recur, surgery time could be extended, but only slightly as nail exchanges take place commonly in a surgery of this type.

Source facts

Action
The recall was initiated with the firm forwarding a recall notice with attached customer response form on January 14, 2010 to the affected consignees. The recall notice informed the consignees of the reason for recall and gave them three steps to follow to inspect the affected Short Trochanteric Nail products. Consignees were provided with product return instruction for any incorrectly bent products found during their inspection. Customers with questions are instructed to contact the firm at (310) 533-9966.
Initiated
2010-01-14
Root cause
Process control
Terminated
2012-04-26
Distribution
Nationwide and Japan
Recall number
Z-1189-2010
Recall status
Terminated
CDRH record ID
88679
FDA classified
2010-03-22
Recall event number
54538

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗