Open original FDA record ↗Recall record
- FDA identifier
- Z-1189-2018
- Company in FDA record
- Zimmer Dental Inc
- FDA classification date
- Mar 5, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two lots of the Tapered Screw-Vent Implants, with different outer thread diameters, were inadvertently commingled during the manufacturing process and may be mislabeled.
Source facts
- Action
- On 11/29/2017, customers were notified via letter. Instructions for distributors include to ensure that affected team members are aware of the recall, immediately locate and quarantine affected devices in inventory, notify consignees if product was distributed, make arrangements to immediately return affected devices, and complete and return the Attachment 1 - Inventory Return Certification Form. Instructions for clinicians include to review the recall notice, examine inventory for affected devices, complete and return the Attachment 2 - Certificate of Acknowledgement, and return all affected devices.
- Initiated
- 2017-11-29
- Root cause
- Packaging process control
- Terminated
- 2019-02-08
- Distribution
- Worldwide distribution US nationwide, Canada, Chile, Costa Rica, Denmark, France, Hungary, Italy, Japan, Lebanon, Portugal, Serbia, Spain, and Taiwan.
- Recall number
- Z-1189-2018
- Recall status
- Terminated
- CDRH record ID
- 160089
- FDA classified
- 2018-03-05
- Recall event number
- 78667