Z-1189-2021: PHYSIO CONTROL LIFEPAK CR2 Defibrillator 3321605-xxx The LIFEPAK CR2 defibrillator is designed for ind

Recall record Z-1189-2021 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1189-2021
Company in FDA record
Physio-Control, Inc.
FDA record creation date
Mar 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints received regarding missing lid magnets which may result in premature battery depletion.

Source facts

Action
On 01/26/2021, the firm, Stryker, sent "URGENT MEDICAL DEVICE SAFETY NOTICE AND CORRECTION" Letters to its customers and distributors via a courier service. The letter describes the product, problem and actions to be taken. Customers were informed that the lid magnet is the primary means by which the device will turn on and off when the lid is opened or closed. If the lid magnet is missing, the device battery can deplete prematurely, even if the device is not powered on. When the magnet is missing, the user can still use the power button to turn the device on and off. The device will automatically turn off within five minutes after being powered on if no patient is detected by the device. The customers were instructed to perform the actions in the notification letter to ensure safe and effective use of their devices. Actions are: 1. Inspect all devices to verify Green Readiness Indicator on device flashes every 6 seconds and the lid magnet is present according to the lid magnet inspection instructions at the end of this notification letter. If any devices are found with Readiness Indicator not flashing or the lid magnet missing, contact Customer Service at 1 800 787 9537 option 2 or by email at medtechsup@stryker.com. 2. Complete and return the Notification Acknowledgment Form attached to the notification letter by Fax at 1 866 448 9567 or by email RSRecall@stryker.com to confirm your receipt of the safety notification. 3. Review the Supplemental Instructions attached to the notification letter and retain the document as supplemental labeling for the device(s). 4. Continue to check device readiness at least monthly in accordance with the Operating Instructions, Maintaining a State of Readiness (pp. 77-78) and the instructions provided herein. Device readiness is indicated by: - All Devices: Green Readiness Indicator on device flashes every 6 seconds. If device is not ready, the Readiness Indicator will not flash. - Devices with Wireless Connectivity: In
Initiated
2021-01-14
Root cause
Component design/selection
Distribution
Worldwide Distribution to U.S.(nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI , SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV and WY O.U.S. countries of: Australia, Austria, Belgium, Canada, Chile, Costa Rica, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Malta, Netherlands, New Zealand, Norway, Poland, Qatar, Saint Barth¿lemy, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Recall number
Z-1189-2021
Recall status
Open, Classified
CDRH record ID
185558
FDA record created
2021-03-09
Recall event number
87217

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗