During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.
Source facts
Action
The firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.
Initiated
2007-04-02
Root cause
Other
Terminated
2008-06-03
Distribution
Product was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).
Recall number
Z-1190-2007
Recall status
Terminated
CDRH record ID
53405
FDA classified
2007-08-23
Recall event number
38324