- FDA identifier
- Z-1190-2012
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Mar 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When subjected to bending, twisting, and malleting, the probe tips break off when used in the sacrum for which they are not indicated for use. These probes are approved for use in the lumbar vertebra.
Source facts
- Action
- An Urgent - Product Safety Notice customer letter was sent via FED EX on 2/6/12 and 2/6/12 to direct consignees in the US identifying the affected product and the issue with it. The letter states that Medtronic Navigation is removing the product from clinical use and will provide a sales credit to customers. In the letters to Medtronic field personnel, it states that local sales and service personnel will be identified to remove the affected product from customer sites and return them for scrapping. Customers are to contact a sales representative if they have any questions regarding the letter.
- Initiated
- 2012-02-06
- Root cause
- Device Design
- Terminated
- 2012-03-26
- Distribution
- Worldwide Distribution -- US and Republic of South Korea.
- Recall number
- Z-1190-2012
- Recall status
- Terminated
- CDRH record ID
- 107459
- FDA classified
- 2012-03-08
- Recall event number
- 61173