Open original FDA record ↗Recall record
- FDA identifier
- Z-1190-2017
- Company in FDA record
- Greatbatch Medical
- FDA record creation date
- Feb 11, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.
Source facts
- Action
- Consignee was sent via e-mail on 10/16/2016 a Greatbatch "Urgent Medical Device Recall" letter dated 10/19/2016 The letter described the purpose of the letter, Reason for Notification, Risk to Health, Recommendation, Instruction to Zimmer. Advised consignee to examine their inventory, quarantine the product and provide a list of the affected countries where the product was distributed. Also requested consignee to complete the Customer Notification Response form and return to Greatbatch at FieldActionCenter@Greatbatch.com. For questions contact your local Greatbatch Medical representative, Monday through Friday via telephone at 1-763-951-8235 8:00 AM (CT) to 5:00 PM (CT), or e-mail us at FieldActionCenter@Greatbatch.com .
- Initiated
- 2016-10-19
- Root cause
- Device Design
- Terminated
- 2019-03-08
- Distribution
- Worldwide Distribution - US Nationwide in the states of IN and the countries of Canada, Austria, Germany, Japan.
- Recall number
- Z-1190-2017
- Recall status
- Terminated
- CDRH record ID
- 152568
- FDA record created
- 2017-02-11
- Recall event number
- 76236