Gas input pressures over 62 psi may cause CPAP pressure oscillation
Source facts
Action
The recalling firm issued a recall letter to their customers dated 9/19/05 informing them of the problem and the corrective action. (fitting unit with external 50 psi pressure regulator for both air and oxygen)
Initiated
2005-09-19
Root cause
Other
Terminated
2006-07-11
Distribution
World wide (including medical facilities in AZ, CA, CO, GA, IA, IL, MA, MD, MN, MO, NC, NE, NH, NJ, OH, OK, OR, PA, PR, TN, TX, VA, WI, and WV. Japan, Ireland, Great Britain, Sweden, Portugal, France, Chile, Argentina, Hong Kong, South Africa, Greece, Egypt, Saudia Arabia, Norway, Spain, Belgium and Canada
Recall number
Z-1191-06
Recall status
Terminated
CDRH record ID
46463
FDA classified
2006-07-01
Recall event number
35608