Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
Source facts
Action
DJO Global sent an Urgent Field Safety Notice letter dated January 10, 2017 to customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to complete the Acknowledgement and receipt Form. Customers were instructed to contact Customer Service at 1-800-456-8696 to explore options for a replacement order.
Initiated
2017-01-10
Root cause
Nonconforming Material/Component
Terminated
2017-03-31
Distribution
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Recall number
Z-1191-2017
Recall status
Terminated
CDRH record ID
152523
FDA classified
2017-02-11
Recall event number
76209