Open original FDA record ↗Recall record
- FDA identifier
- Z-1191-2023
- Company in FDA record
- Stryker GmbH
- FDA classification date
- Mar 1, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.
Source facts
- Action
- The firm issue an URGENT MEDICAL DEVICE RECALLL its consignees by letter on 02/07/2023. The notice explained the product and the issue and requested the following actions be taken: 1. Check inventory for affected product and remove from use 2, Sign and return the enclosed Business Reply Form by email to fieldaction@stryker.com to confirm receipt of this notification/documenting product disposition. Stryker will will contact you to arrange for the return of the affected device(s) 3. Spread awareness of this communication internally until all required actions have been completed within your facility. 4. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so that Stryker can inform the recipients appropriately.
- Initiated
- 2023-02-07
- Root cause
- Process change control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1191-2023
- Recall status
- Open, Classified
- CDRH record ID
- 198463
- FDA classified
- 2023-03-01
- Recall event number
- 91666