Open original FDA record ↗Recall record
- FDA identifier
- Z-1192-2019
- Company in FDA record
- Medrobotics Corporation
- FDA record creation date
- Apr 17, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The weld may break, resulting in the rod separating from the cone of the long and short cone Obturators.
Source facts
- Action
- On January 25, 2019, the firm issued an Urgent Voluntary Medical Device Recall notification letter to its customers. The letter informed customers of the potential for the handle of the obturator to detach with the cone remaining in the patient's rectum. Patients were asked to take the following actions: 1. Immediately check your inventory and quarantine all Obturators that you have on hand to prevent their further use. Note: An Obturator may be located in the same sterilization tray as the Flex Rectoscope. 2. Please immediately complete the attached Acknowledgement form -- even if you have no Obturators -and return the form it to Medrobotics by fax: +1-508-823-1703 or by email customerservice@medrobotics.com. 3. Ensure relevant staff members -- including clinicians, OR staff, engineers, and product returns staff -- are informed of this recall. 4. If you have supplied any Obturators to any other organization, please advise that organization of this recall and notify Med robotics of that organization's name, address, and contact person so that we may contact them immediately. Please contact Med robotics Customer Service at 1 +508-692-6104 or at customerservice@medrobotics.com if you have any questions or need additional information.
- Initiated
- 2019-01-25
- Root cause
- Process control
- Terminated
- 2020-12-11
- Distribution
- Worldwide Distribution - US Nationwide in the states of CO, DC, FL, IL, MD, MO, NY, OK, PA, TX, and VA. Foreign distribution to Australia and Italy.
- Recall number
- Z-1192-2019
- Recall status
- Terminated
- CDRH record ID
- 171553
- FDA record created
- 2019-04-17
- Recall event number
- 82403